YAO Xueqi, DONG Yunning, YAN Rui, CHEN Junru, DING Shuang, KANG Pengwei, LU Yaowen. Performance Study on Portable Calibration Apparatus of Mass Spectrometers for Partial Pressure[J]. CHINESE JOURNAL OF VACUUM SCIENCE AND TECHNOLOGY, 2024, 44(4): 306-311. DOI: 10.13922/j.cnki.cjvst.202401010
Citation: YAO Xueqi, DONG Yunning, YAN Rui, CHEN Junru, DING Shuang, KANG Pengwei, LU Yaowen. Performance Study on Portable Calibration Apparatus of Mass Spectrometers for Partial Pressure[J]. CHINESE JOURNAL OF VACUUM SCIENCE AND TECHNOLOGY, 2024, 44(4): 306-311. DOI: 10.13922/j.cnki.cjvst.202401010

Performance Study on Portable Calibration Apparatus of Mass Spectrometers for Partial Pressure

  • This paper tested part performance of the latest developed portable calibration apparatus of mass spectrometers for partial pressure. The ultimate vacuum of the device was measured using a reference ionization gauge; The diameter and the thickness of the gas-operated element which conductance is about 10−2 m3/s could be measured accurately and the conductance was calculated using formular under the molecular flow condition. The conductance of the gas-intaked conductance elements with micro-meter apertures was measured using constant volume experiments. The back-streaming ratio is obtained using the same vacuum gauge to measure the gas pressure at both sides of the gas-operated conductance elements. The experimental results proved that the ultimate vacuum of the device is 4.8×10−7 Pa. The conductance of the gas-exhausted element for N2 is 1.37×10−2 m3/s. The conductance of two gas-intaked conductance elements for N2 are 3.81×10−6 m3/s and 4.50×10−8 m3/s respectively. The back-streaming ratio for N2 is 0.193 .Based on analysis, the calibration range of the device for N2 is (4×10−2~5×10−6)Pa.
  • loading

Catalog

    Turn off MathJax
    Article Contents

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return
    Return